ISO/IEC 17025 method validation and verification

Every accredited result rests on a method the laboratory has shown to be fit for purpose. This guide explains the difference between validation and verification, what clause 7.2 asks for, and how to document it so an assessor can follow it.

Last updated:

In short

Under ISO/IEC 17025:2017, a laboratory verifies standard methods (clause 7.2.1.5): it shows that it can perform a published, already validated method properly and achieve the required performance before reporting results. It validates methods (clause 7.2.2) when they are non-standard, developed by the laboratory, or standard methods used outside their intended scope or otherwise modified. Validation determines the performance characteristics relevant to the customer's needs — such as measurement range, accuracy, bias, repeatability and reproducibility, measurement uncertainty, limits of detection and quantification, selectivity, linearity and robustness — and ends with a documented statement that the method is fit for its intended use. Records must show the procedure used, the requirements, the characteristics determined, the results and that statement.

Validation or verification — which one applies?

The distinction decides how much work a new method costs. Verification asks "can we do this correctly?" for a method someone else has already validated. Validation asks "does this method do what it is supposed to?" and has to be answered by the laboratory itself.

SituationWhat is required
Published standard method (ISO, EN, ASTM, IEC, DIN…) used as writtenVerification (7.2.1.5): show that you achieve the performance stated in the standard with your staff, equipment and environment.
New revision of a standard methodVerification of the changes — review what changed and re-verify the affected performance characteristics.
Laboratory-developed or in-house methodFull validation (7.2.2.1).
Standard method outside its scope (other material, range, temperature, sample type)Validation of the extension — at least for the characteristics the change can affect.
Standard method with modifications (shortened test time, different fixture, other instrument principle)Validation of the modification, and assessment of its influence on results (7.2.2.2).
Customer-specified or manufacturer methodTreated as non-standard unless it is published and validated — validation, with the customer's agreement on the requirements.

In test laboratories for products — electrical, mechanical, environmental, battery or EMC testing — most methods are standard methods, so verification is the everyday case. Validation comes up mainly with customer test specifications, accelerated or shortened procedures and in-house fixtures.

Performance characteristics to determine

Clause 7.2.2.3 requires that the performance characteristics of a validated method are relevant to the customer's needs and consistent with specified requirements. The standard names typical characteristics in a note; not all apply to every method. Choose and justify the ones that matter.

CharacteristicQuestion it answers
Measurement rangeOver which range of values does the method give valid results?
Accuracy / trueness (bias)How far is the mean result from a reference value, e.g. a certified reference material or a reference method?
Precision: repeatabilityHow much do results scatter with the same operator, equipment and short time interval?
Precision: intermediate precision / reproducibilityHow much do results scatter across operators, days, instruments — or across laboratories?
Measurement uncertaintyWhich interval around the result contains the true value with a stated confidence?
Limit of detection / quantificationWhat is the smallest value that can be detected or quantified reliably?
Selectivity / cross-sensitivityDo other components, the matrix or interference affect the result?
LinearityIs the response proportional over the range, or does it need correction?
Robustness (ruggedness)How sensitive are results to small changes in conditions — temperature, fixture torque, timing, ramp rates?

A validation plan in seven steps

  1. 1

    Define the requirements

    Write down what the method has to achieve: measurand, range, required uncertainty or tolerance, the decision the results will support. Agree them with the customer where relevant — this is the yardstick for "fit for purpose".

  2. 2

    Select the characteristics

    Pick the performance characteristics that matter for these requirements and justify the ones you leave out.

  3. 3

    Set acceptance criteria in advance

    Define before testing what counts as a pass: e.g. bias within ±1 %, repeatability standard deviation below a set value. Criteria defined afterwards are an assessment finding waiting to happen.

  4. 4

    Choose the techniques

    The standard lists several in a note to 7.2.2.1: reference standards or reference materials, systematic assessment of influence factors, robustness tests by varying controlled parameters, comparison with another validated method, interlaboratory comparisons, and uncertainty evaluation based on the method's principles.

  5. 5

    Run the experiments

    With calibrated equipment, authorised staff and recorded conditions — the validation data are technical records like any other.

  6. 6

    Evaluate and decide

    Compare results against the acceptance criteria and write the statement on validity: the method is (or is not) fit for the intended use, within stated limits.

  7. 7

    Release and monitor

    Release the method version, authorise staff, and monitor ongoing validity with quality control and proficiency testing (clause 7.7). Re-validate when the method changes.

What the validation record must contain

Clause 7.2.2.4 lists the minimum content of the retained records:

  • The validation procedure used.
  • The specification of the requirements.
  • The determination of the performance characteristics of the method.
  • The results obtained.
  • A statement on the validity of the method, detailing its fitness for the intended use.

For verification, the standard is less specific, but assessors expect the same logic on a smaller scale: which performance the standard method claims, what you measured, and the conclusion that you meet it. Keep the raw data traceable — which items, which instruments, which calibration status, which operator.

Typical assessment findings

  • A standard method was introduced without any verification record, or verification consists only of "method works".
  • A modified standard method (shorter dwell time, other fixture) is treated as the standard method and was never validated.
  • Acceptance criteria were defined after the results were known.
  • The validation report has no statement on fitness for intended use.
  • A new revision of the standard was adopted without reviewing what changed.
  • Validation data cannot be traced to the instruments and their calibration status at the time.
  • Staff perform the method without an authorisation that references the released method version.

Several of these come down to version control: which method revision was released when, and which test jobs were run with which revision. When method versions, test sequences, instruments and results are linked in one system, that question is answered by a query instead of a search through folders.

Frequently asked questions

What is the difference between method validation and method verification?

Validation provides objective evidence that a method is fit for its intended use — it establishes the method's performance characteristics. Verification confirms that a laboratory can correctly perform a method that has already been validated, typically a published standard method, and achieve its stated performance.

Do standard methods have to be validated under ISO/IEC 17025?

Not if they are used as published and within their scope — then verification is sufficient (clause 7.2.1.5). Validation is required once a standard method is modified or used outside its intended scope (clause 7.2.2.1).

How many repetitions does a validation need?

ISO/IEC 17025 does not specify a number. It depends on the characteristic and the statistical confidence needed. Many laboratories use at least 10 repetitions for repeatability and several series on different days or with different operators for intermediate precision; sector guides and the accreditation body may give more specific expectations.

When does a method have to be re-validated?

When changes are made to a validated method, the laboratory must determine their influence and, where they affect the original validation, perform a new validation (clause 7.2.2.2). Typical triggers are new equipment principles, new fixtures, extended ranges or new sample types.

Know which method version produced which result

TestLabIQ links test sequences, instruments with calibration status, measurement data and reports — so validation evidence and routine results stay traceable.